Contact Angle & Surface Tension Tensiometry & Surface/Interfacial Tension Measurement

MBP – Bubble Pressure Tensiometer

Part Number: MBP
The MBP 200 by DataPhysics Instruments employs the maximum bubble pressure method to accurately measure dynamic surface tension of liquid solutions. Gas is directed through a capillary into the liquid; the internal pressure of the forming bubble is recorded at its peak, which corresponds to distinct bubble age and pressure values. This enables dynamic tension readings across a broad range of 5 ms to 200 s, and 10–100 mN/m, suitable for surfactants, oils, and more. With optional temperature control (-15 °C to 135 °C) and dosing accessories, it offers comprehensive surface chemistry analysis in labs or production environments.
Wide Dynamic Range: 10–100 mN/m in 5 ms–200 s
Captures dynamic surface tension from rapid bubble formation to equilibrium, empowering users to observe surfactant kinetics and fluid aging behaviors.
High-Pressure Capability up to 3,400 Pa
Enables measurements of highly viscous liquids and industrial oils, thanks to robust pressure sensing.
Temperature-Controlled Sampling from –15 °C to 135 °C
Optional Peltier or liquid circulator systems allow dynamic tension testing under varied thermal conditions.
Flexible Capillary Options
Supports both reusable glass and disposable plastic capillaries, minimizing cleaning effort and reducing cross-contamination.
Automated Dosing and Concentration Series
With the LDU 25 liquid dosing unit, enables automated surfactant concentration sweeps, visualizing concentration-dependent tension behavior.
Safety Features to Protect Capillaries
Splash guard and collision protection ensure long capillary life and device reliability.
Specification Value
Dynamic Surface Tension Range 10 – 100 mN/m
Surface Age Range 5 ms – 200 s
Pressure Measurement Resolution Up to 3,400 Pa differential
Sampling Rate (Bubble Pressure) 25,000 Hz
Sample Stage Travel 105 mm range, 46 nm/s – 12 mm/s resolution 24 nm
Temperature Range –15 °C to 135 °C (Peltier); –10 °C to 130 °C (liquid circulator)
Temperature Sensors 2 × Pt100 inputs (–60 °C to +450 °C, ±0.01 K)
Weight & Dimensions 22 kg; 360 × 230 × 565 mm
Power Supply 100–240 VAC, 50/60 Hz, 70 W
Control & Interface TP 50 touchscreen optional; PC control via MBPSoftware
Software Functions Single/sweep ages, concentration series, adsorption/diffusion coefficients
Pdf